Recall Z-1047-2025— ROi CPS LLC

Class I · Ongoing

Class I FDA medical device recall by ROi CPS LLC, initiated 2024-12-26. FDA's recall record names no specific 510(k) clearance for this event. Product: regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.. Reason: There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient...

Recalling firm
ROi CPS LLC
Classification
Class I
Status
Ongoing
Initiated
2024-12-26
Firm location
Republic, MO, United States

Product description

regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.

Reason for recall

There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..