Recall Z-1045-2023— MEDLINE INDUSTRIES, LP - Northfield

Class II · Ongoing

Class II FDA medical device recall by MEDLINE INDUSTRIES, LP - Northfield, initiated 2022-12-22. It names one FDA 510(k) clearance: K181280. Product: MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008. Reason: The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque..

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Classification
Class II
Status
Ongoing
Initiated
2022-12-22
Firm location
Northfield, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008

Reason for recall

The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.