Recall Z-1045-2013— Zimmer, Inc.
Class II · Terminated
Class II FDA medical device recall by Zimmer, Inc., initiated 2013-02-08, terminated 2014-12-09. It names 2 FDA 510(k) clearances: K061345, K053267. Product: Zimmer NexPosure, retractor portals of the NexPosure MIS Access System, Part Numbers or REF numbers: N1830R80-100, N1830R80-090¿, N1830R80-080¿, N1830R80-070¿, N1830R80-060, N1830R50-100, N1830R50-090, N1830R50-080, ¿ N1830R50-070, N1830R50-060, N1830L80-100, N1830L80-090, N1830L80-080, N1830L80-070, N1830L80-060, N1830L50-100, N1830L50-090, N1830L50-080, N1830L50-070, N1830L50-060, ¿ N182450-085, N182450-075, N182450-065, N182450-055, N182450-045.¿ Retractor access system that allows placement of posterior spinal fixation implants.. Reason: Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased risk of contamination (i.e.; loss of product sterility) if proper aseptic technique is not followed when transferring the inner package into the sterile field. Patient infection may o.
- Recalling firm
- Zimmer, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-02-08
- Terminated
- 2014-12-09
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Zimmer NexPosure, retractor portals of the NexPosure MIS Access System, Part Numbers or REF numbers: N1830R80-100, N1830R80-090¿, N1830R80-080¿, N1830R80-070¿, N1830R80-060, N1830R50-100, N1830R50-090, N1830R50-080, ¿ N1830R50-070, N1830R50-060, N1830L80-100, N1830L80-090, N1830L80-080, N1830L80-070, N1830L80-060, N1830L50-100, N1830L50-090, N1830L50-080, N1830L50-070, N1830L50-060, ¿ N182450-085, N182450-075, N182450-065, N182450-055, N182450-045.¿ Retractor access system that allows placement of posterior spinal fixation implants.
Reason for recall
Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased risk of contamination (i.e.; loss of product sterility) if proper aseptic technique is not followed when transferring the inner package into the sterile field. Patient infection may o