Recall Z-1044-2015— Philips Medical Systems, Inc.
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems, Inc., initiated 2013-10-14, terminated 2017-01-23. It names one FDA 510(k) clearance: K023441. Product: Philips Multi Diagnost Eleva II. Reason: The possibility of the Reference Air Kerma Rate for low dose mode will not be reduced to 50% of the value for the normal mode, but will be approximately 70-80%, leading to a higher patient dose than expected..
- Recalling firm
- Philips Medical Systems, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-10-14
- Terminated
- 2017-01-23
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips Multi Diagnost Eleva II
Reason for recall
The possibility of the Reference Air Kerma Rate for low dose mode will not be reduced to 50% of the value for the normal mode, but will be approximately 70-80%, leading to a higher patient dose than expected.