Recall Z-1042-2024— BALT USA, LLC
Class II · Ongoing
Class II FDA medical device recall by BALT USA, LLC, initiated 2024-01-03. It names one FDA 510(k) clearance: K220699. Product: The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to embolize peripheral vascular abnormalities. The devices are to be placed to create blood stasis, reducing flow into the anomaly, and thrombosing the target site. The Prestige Coil System is comprised of the implant coil and a pusher assembly. There are a total of 81 Prestige coil models. Vascular Embolization Device Prestige Coil Model Numbers and Product description: PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES1137COM18 Prestige Coil 11mm x 37cm Complex-18 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack PRES0930CPKPL Prestige Coil 9mm x 30cm Prestige Plus Complex Pack PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack PRES0830HPKPL Prestige Coil 8mm x 30cm Prestige Plus Helical Pack PRES0930HPKPL Prestige Coil 9mm x 30cm Prestige Plus Helical Pack PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18 PRES0204CXPPL Prestige Coil 2mm x 4cm Prestige Plus Complex Xtra Pac PRES0206CXPPL Prestige Coil 2mm x 6cm Prestige Plus Complex Xtra Pac PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack PRES0210CXP10 Prestige Coil 2mm x 10cm Complex-10 Xtra Pack PRES0258CXP10 Prestige Coil 2.5mm x 8cm Complex-10 Xtra Pack PRES0417CPK10 Prestige Coil 4mm x 17cm Complex-10 Xtra Pack PRES0520CPK10 Prestige Coil 5mm x 20cm Complex-10 Xtra Pack PRES0615CPK10 Prestige Coil 6mm x 15cm Complex-10 Xtra Pack PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES2525HXPPL Prestige Coil 2.5mm x 2.5cm Prestige Plus Helical Xtra Pac PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CRG10 Prestige Coil 4mm x 13cm Complex 10 Regular PRES0153CXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Complex Xtra Pac PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0306HPK10 Prestige Coil 3mm x 6cm Helical-10 Pack PRES0310HPK10 Prestige Coil 3mm x 10cm Helical-10 Pack PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack PRES0930H. Reason: Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli..
- Recalling firm
- BALT USA, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-01-03
- Firm location
- Irvine, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to embolize peripheral vascular abnormalities. The devices are to be placed to create blood stasis, reducing flow into the anomaly, and thrombosing the target site. The Prestige Coil System is comprised of the implant coil and a pusher assembly. There are a total of 81 Prestige coil models. Vascular Embolization Device Prestige Coil Model Numbers and Product description: PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES1137COM18 Prestige Coil 11mm x 37cm Complex-18 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack PRES0930CPKPL Prestige Coil 9mm x 30cm Prestige Plus Complex Pack PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack PRES0830HPKPL Prestige Coil 8mm x 30cm Prestige Plus Helical Pack PRES0930HPKPL Prestige Coil 9mm x 30cm Prestige Plus Helical Pack PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18 PRES0204CXPPL Prestige Coil 2mm x 4cm Prestige Plus Complex Xtra Pac PRES0206CXPPL Prestige Coil 2mm x 6cm Prestige Plus Complex Xtra Pac PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack PRES0210CXP10 Prestige Coil 2mm x 10cm Complex-10 Xtra Pack PRES0258CXP10 Prestige Coil 2.5mm x 8cm Complex-10 Xtra Pack PRES0417CPK10 Prestige Coil 4mm x 17cm Complex-10 Xtra Pack PRES0520CPK10 Prestige Coil 5mm x 20cm Complex-10 Xtra Pack PRES0615CPK10 Prestige Coil 6mm x 15cm Complex-10 Xtra Pack PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES2525HXPPL Prestige Coil 2.5mm x 2.5cm Prestige Plus Helical Xtra Pac PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CRG10 Prestige Coil 4mm x 13cm Complex 10 Regular PRES0153CXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Complex Xtra Pac PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0306HPK10 Prestige Coil 3mm x 6cm Helical-10 Pack PRES0310HPK10 Prestige Coil 3mm x 10cm Helical-10 Pack PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack PRES0930H
Reason for recall
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.