Recall Z-1039-2021— Aomori Olympus Co., Ltd.

Class II · Terminated

Class II FDA medical device recall by Aomori Olympus Co., Ltd., initiated 2021-01-04, terminated 2024-03-11. It names one FDA 510(k) clearance: K153625. Product: NEEDLEMASTER 6MMx 25G UPPER, model no. NM-610L-0625 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.. Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used..

Recalling firm
Aomori Olympus Co., Ltd.
Classification
Class II
Status
Terminated
Initiated
2021-01-04
Terminated
2024-03-11
Firm location
Kuroishi, N/A, Japan

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

NEEDLEMASTER 6MMx 25G UPPER, model no. NM-610L-0625 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.

Reason for recall

Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.