Recall Z-1036-2017— Olympus Scientific Solutions Americas

Class II · Terminated

Class II FDA medical device recall by Olympus Scientific Solutions Americas, initiated 2017-01-07, terminated 2019-08-27. FDA's recall record names no specific 510(k) clearance for this event. Product: Olympus Scientific Solutions Americas Corporation (OSSA) VANTA¿ XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems. Reason: It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on..

Recalling firm
Olympus Scientific Solutions Americas
Classification
Class II
Status
Terminated
Initiated
2017-01-07
Terminated
2019-08-27
Firm location
Waltham, MA, United States

Product description

Olympus Scientific Solutions Americas Corporation (OSSA) VANTA¿ XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems

Reason for recall

It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.