Recall Z-1035-2015— GE Healthcare
Class I · Terminated
Class I FDA medical device recall by GE Healthcare, initiated 2015-01-07, terminated 2015-09-10. It names 5 FDA 510(k) clearances: K041476, K040444, K030874 and 2 more. Product: This correction applies to all GE Healthcare MRI systems with superconducting magnets: Discovery MR450, Discovery MR750, Signa HDx, Signa HDxt, Signa HDi, GE 1.5T Signa HDe, Optima MR360, Brivo MR355, Signa Excite-HD 1.5T & 3T, Signa Excite 1.5T HD Twinspeed, Signa Excite 1.5T HD Echospeed, Signa Excite 1.5T HD Highspeed, 1.5T Signa Infinity TwinSpeed , 1.5T Signa Infinity EchoSpeed Plus, 1.5T Signa Infinity HiSpeed Plus, 1.0T Signa Infinity HiSpeed Plus. 1.0T Signa Infinity SmartSpeed, Signa EXCITE 3.0T, Signa EXCITE 3.0T HD, Signa Excite 1.5T TwinSpeed, Signa Excite 1.5T EchoSpeed, Signa Excite 1.5T HiSpeed, Signa Excite 1.5T SmartSpeed, Signa Excite 1.5T, Signa Excite 3.0T, Signa Contour/I, Signa OpenSpeed, Vectra, MR Max, Optima MR450w GEM, Discovery MR750W GEM, Optima MR450w. Product Usage: Magnetic resonance (MR) diagnostic devices are intended for producing images of the internal structures of the body based on the spatial distribution of molecules exhibiting magnetic resonance. Most MR systems are intended for diagnostic use.. Reason: At certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times..
- Recalling firm
- GE Healthcare
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2015-01-07
- Terminated
- 2015-09-10
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
5 on FDA's record- K041476 — GE SIGNA EXCITE 1.5T MR SYSTEM, GE SIGNA EXCITE 3.0T MR SYSTEM
- K040444 — GE SIGNA 3.0T WITH EXCITE MR SYSTEM
- K030874 — GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM
- K013636 — SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY
- K012200 — 1.5T INFINITY TWINSPEED MR SYSTEM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
This correction applies to all GE Healthcare MRI systems with superconducting magnets: Discovery MR450, Discovery MR750, Signa HDx, Signa HDxt, Signa HDi, GE 1.5T Signa HDe, Optima MR360, Brivo MR355, Signa Excite-HD 1.5T & 3T, Signa Excite 1.5T HD Twinspeed, Signa Excite 1.5T HD Echospeed, Signa Excite 1.5T HD Highspeed, 1.5T Signa Infinity TwinSpeed , 1.5T Signa Infinity EchoSpeed Plus, 1.5T Signa Infinity HiSpeed Plus, 1.0T Signa Infinity HiSpeed Plus. 1.0T Signa Infinity SmartSpeed, Signa EXCITE 3.0T, Signa EXCITE 3.0T HD, Signa Excite 1.5T TwinSpeed, Signa Excite 1.5T EchoSpeed, Signa Excite 1.5T HiSpeed, Signa Excite 1.5T SmartSpeed, Signa Excite 1.5T, Signa Excite 3.0T, Signa Contour/I, Signa OpenSpeed, Vectra, MR Max, Optima MR450w GEM, Discovery MR750W GEM, Optima MR450w. Product Usage: Magnetic resonance (MR) diagnostic devices are intended for producing images of the internal structures of the body based on the spatial distribution of molecules exhibiting magnetic resonance. Most MR systems are intended for diagnostic use.
Reason for recall
At certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.
Data sourced from openFDA. This site is unofficial and independent of the FDA.