Recall Z-1035-2013— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2013-02-12, terminated 2013-12-23. It names one FDA 510(k) clearance: K060256. Product: UniCel DxC 600i Synchron Access Clinical System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.. Reason: Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2013-02-12
Terminated
2013-12-23
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

UniCel DxC 600i Synchron Access Clinical System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.

Reason for recall

Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results.