Recall Z-1034-2026— IMRIS Imaging Inc
Class II · Ongoing
Class II FDA medical device recall by IMRIS Imaging Inc, initiated 2025-12-08. It names one FDA 510(k) clearance: K103493. Product: Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number: 119695-000, 119629-000, 119630-000, 113803-000, 113803-600, 113802-000, 113802-600 Software Version: Not Applicable Product Description: The HFD100 Head Fixation Device provides rigid skeletal fixation within the optimal imaging envelope while introducing minimal artifacts in the acquired images. There are two different assemblies for the HFD100. The standard HFD100 is equipped with a table adapter for use with either the operating room table or the angiography table. For sites with a Hill-Rom TruSystem 7500 Hybrid Neuro MR Operating Table (ORT400), the HFD100 has a split block clamp assembly (Trumpf Mount Clamp Assembly) which connects directly to the table back section. The HFD100 system is also comprised of the linkage components to mount the skull clamp to the table adapter, and the 3-pin skull clamp. The linkage system is used to position the skull clamp in the necessary position, without colliding with the magnet bore. Component: Not a component.. Reason: Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury..
- Recalling firm
- IMRIS Imaging Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-08
- Firm location
- Chaska, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number: 119695-000, 119629-000, 119630-000, 113803-000, 113803-600, 113802-000, 113802-600 Software Version: Not Applicable Product Description: The HFD100 Head Fixation Device provides rigid skeletal fixation within the optimal imaging envelope while introducing minimal artifacts in the acquired images. There are two different assemblies for the HFD100. The standard HFD100 is equipped with a table adapter for use with either the operating room table or the angiography table. For sites with a Hill-Rom TruSystem 7500 Hybrid Neuro MR Operating Table (ORT400), the HFD100 has a split block clamp assembly (Trumpf Mount Clamp Assembly) which connects directly to the table back section. The HFD100 system is also comprised of the linkage components to mount the skull clamp to the table adapter, and the 3-pin skull clamp. The linkage system is used to position the skull clamp in the necessary position, without colliding with the magnet bore. Component: Not a component.
Reason for recall
Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury.