Recall Z-1034-2022— Olympus Corporation of the Americas
Class II · Ongoing
Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2022-03-18. It names one FDA 510(k) clearance: K011314. Product: Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.. Reason: GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients.
- Recalling firm
- Olympus Corporation of the Americas
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-03-18
- Firm location
- Center Valley, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.
Reason for recall
GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients