Recall Z-1032-2026— Securitas Healthcare LLC

Class II · Ongoing

Class II FDA medical device recall by Securitas Healthcare LLC, initiated 2025-12-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3. Reason: Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing..

Recalling firm
Securitas Healthcare LLC
Classification
Class II
Status
Ongoing
Initiated
2025-12-08
Firm location
Lincoln, NE, United States

Product description

Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3

Reason for recall

Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.