Recall Z-1031-2026— Securitas Healthcare LLC
Class II · Ongoing
Class II FDA medical device recall by Securitas Healthcare LLC, initiated 2025-12-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3. Reason: Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing..
- Recalling firm
- Securitas Healthcare LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-08
- Firm location
- Lincoln, NE, United States
Product description
Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3
Reason for recall
Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.