Recall Z-1028-2013— Roche Diagnostics Operations, Inc.
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2013-03-01, terminated 2014-06-04. It names one FDA 510(k) clearance: K110923. Product: Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas¿ CT/NG 4800 System, for in vitro diagnostics.. Reason: Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent..
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-03-01
- Terminated
- 2014-06-04
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas¿ CT/NG 4800 System, for in vitro diagnostics.
Reason for recall
Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent.