Recall Z-1026-2015— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2014-12-15, terminated 2016-02-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Beckman Coulter, Lactate Assay, Catalog No. OSR6X93. Reason: N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC interference may lead to falsely low results..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2014-12-15
Terminated
2016-02-19
Firm location
Brea, CA, United States

Product description

Beckman Coulter, Lactate Assay, Catalog No. OSR6X93

Reason for recall

N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC interference may lead to falsely low results.