Recall Z-1022-2016— Philips Medical Systems (Cleveland) Inc
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2015-12-17, terminated 2018-07-02. It names one FDA 510(k) clearance: K080927. Product: BrightView XCT & BrightView XCT Upgrade, gamma cameras; For Single Proton Emission Computed Tomography.. Reason: The firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of the patient pallet. If this occurs, collision sensors would pause the scan. Issue 2: During a scan, the gantry motion may halt prematurely. This cause.
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-12-17
- Terminated
- 2018-07-02
- Firm location
- Cleveland, OH, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BrightView XCT & BrightView XCT Upgrade, gamma cameras; For Single Proton Emission Computed Tomography.
Reason for recall
The firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of the patient pallet. If this occurs, collision sensors would pause the scan. Issue 2: During a scan, the gantry motion may halt prematurely. This cause
Data sourced from openFDA. This site is unofficial and independent of the FDA.