Recall Z-1020-2023— SunMed Holdings, LLC

Class II · Ongoing

Class II FDA medical device recall by SunMed Holdings, LLC, initiated 2022-12-22. FDA's recall record names no specific 510(k) clearance for this event. Product: AirQ3 Self Pressurizing ILA with Automatic Cuff Inflation and GA Size REF Number 0.0 60005 0.5 60055 1.0 60105 1.5 60155 2.0 60205 3.0 60305 4.0 60405. Reason: There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter..

Recalling firm
SunMed Holdings, LLC
Classification
Class II
Status
Ongoing
Initiated
2022-12-22
Firm location
Grand Rapids, MI, United States

Product description

AirQ3 Self Pressurizing ILA with Automatic Cuff Inflation and GA Size REF Number 0.0 60005 0.5 60055 1.0 60105 1.5 60155 2.0 60205 3.0 60305 4.0 60405

Reason for recall

There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.

Data sourced from openFDA. This site is unofficial and independent of the FDA.