Recall Z-1018-2026— Canon Medical System, USA, INC.

Class II · Ongoing

Class II FDA medical device recall by Canon Medical System, USA, INC., initiated 2025-12-19. It names one FDA 510(k) clearance: K191128. Product: VANTAGE TITAN 3T Model MRT-3010/MEXL-3010. Reason: There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room..

Recalling firm
Canon Medical System, USA, INC.
Classification
Class II
Status
Ongoing
Initiated
2025-12-19
Firm location
Tustin, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

VANTAGE TITAN 3T Model MRT-3010/MEXL-3010

Reason for recall

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.