Recall Z-1018-2022— Becton Dickinson & Co.

Class II · Terminated

Class II FDA medical device recall by Becton Dickinson & Co., initiated 2022-03-09, terminated 2024-01-19. FDA's recall record names no specific 510(k) clearance for this event. Product: BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10. Reason: When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes..

Recalling firm
Becton Dickinson & Co.
Classification
Class II
Status
Terminated
Initiated
2022-03-09
Terminated
2024-01-19
Firm location
Sparks, MD, United States

Product description

BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

Reason for recall

When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.