Recall Z-1018-2019— Randox Laboratories Ltd.
Class II · Ongoing
Class II FDA medical device recall by Randox Laboratories Ltd., initiated 2018-12-03. It names one FDA 510(k) clearance: K023763. Product: Fructosamine Control 3, FR2996 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.. Reason: The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and controls has now been re-assigned by a decrease of 25%..
- Recalling firm
- Randox Laboratories Ltd.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2018-12-03
- Firm location
- Crumlin (North), N/A, Ireland
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Fructosamine Control 3, FR2996 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.
Reason for recall
The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and controls has now been re-assigned by a decrease of 25%.
Data sourced from openFDA. This site is unofficial and independent of the FDA.