Recall Z-1017-2015— Covidien

Class I · Terminated

Class I FDA medical device recall by Covidien, initiated 2014-12-10, terminated 2016-03-31. It names one FDA 510(k) clearance: K071664. Product: Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01. Sterile EO. Product Usage: Intended for controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.. Reason: A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal balloon inflation port incorrectly labeled as proximal, and, the proximal balloon port incorrectly labeled as distal..

Recalling firm
Covidien
Classification
Class I
Status
Terminated
Initiated
2014-12-10
Terminated
2016-03-31
Firm location
Plymouth, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01. Sterile EO. Product Usage: Intended for controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

Reason for recall

A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal balloon inflation port incorrectly labeled as proximal, and, the proximal balloon port incorrectly labeled as distal.