Recall Z-1013-2026— Medline Industries, LP
Class I · Ongoing
Class I FDA medical device recall by Medline Industries, LP, initiated 2025-11-26. FDA's recall record names no specific 510(k) clearance for this event. Product: MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4.. Reason: Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire..
- Recalling firm
- Medline Industries, LP
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2025-11-26
- Firm location
- Northfield, IL, United States
Product description
MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4.
Reason for recall
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Data sourced from openFDA. This site is unofficial and independent of the FDA.