Recall Z-1012-2026— Medline Industries, LP

Class I · Ongoing

Class I FDA medical device recall by Medline Industries, LP, initiated 2025-11-26. FDA's recall record names no specific 510(k) clearance for this event. Product: MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L. Reason: Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire..

Recalling firm
Medline Industries, LP
Classification
Class I
Status
Ongoing
Initiated
2025-11-26
Firm location
Northfield, IL, United States

Product description

MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L

Reason for recall

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Data sourced from openFDA. This site is unofficial and independent of the FDA.