Recall Z-1010-2018— Fresenius Medical Care Renal Therapies Group, LLC
Class II · Terminated
Class II FDA medical device recall by Fresenius Medical Care Renal Therapies Group, LLC, initiated 2017-09-29, terminated 2018-08-09. It names one FDA 510(k) clearance: K070177. Product: NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.. Reason: The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L..
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-09-29
- Terminated
- 2018-08-09
- Firm location
- Waltham, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Reason for recall
The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.