Recall Z-1009-2023— Abbott Molecular, Inc.
Class II · Ongoing
Class II FDA medical device recall by Abbott Molecular, Inc., initiated 2022-12-09. FDA's recall record names no specific 510(k) clearance for this event. Product: Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.. Reason: The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured..
- Recalling firm
- Abbott Molecular, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-12-09
- Firm location
- Des Plaines, IL, United States
Product description
Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.
Reason for recall
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.
Data sourced from openFDA. This site is unofficial and independent of the FDA.