Recall Z-1008-2024— Brasseler USA I Lp

Class II · Ongoing

Class II FDA medical device recall by Brasseler USA I Lp, initiated 2024-01-11. FDA's recall record names no specific 510(k) clearance for this event. Product: Brasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023, REF 001156T0), carbide dental bur, 10 burs/package, Rx.. Reason: The device has a grip detail (right angle latch) out of specification causing the bur to not adequately latch in the handpiece..

Recalling firm
Brasseler USA I Lp
Classification
Class II
Status
Ongoing
Initiated
2024-01-11
Firm location
Savannah, GA, United States

Product description

Brasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023, REF 001156T0), carbide dental bur, 10 burs/package, Rx.

Reason for recall

The device has a grip detail (right angle latch) out of specification causing the bur to not adequately latch in the handpiece.

Data sourced from openFDA. This site is unofficial and independent of the FDA.