Recall Z-1006-2014— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2013-10-11, terminated 2017-04-14. FDA's recall record names no specific 510(k) clearance for this event. Product: VERSA-FX KEYLESS 90D X 8 and X 11, VERSA-FX KEYLESS 95D X 8, X 11, and X 14, VERSA-FX KEYLESS 140D X 4 and X 5, VERSA-FX KEYLESS 135D X 4 and X 5, VERSA-FX SHORT TUBE, 130D, 135D and 140D, VERSA-FX STANDARD TUBE, 1 and VERSA-FX SPRCNDLR TUBE, 9 Item codes starting with 001980, 001981, 0011990 and 0011991. Reason: Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-10-11
Terminated
2017-04-14
Firm location
Warsaw, IN, United States

Product description

VERSA-FX KEYLESS 90D X 8 and X 11, VERSA-FX KEYLESS 95D X 8, X 11, and X 14, VERSA-FX KEYLESS 140D X 4 and X 5, VERSA-FX KEYLESS 135D X 4 and X 5, VERSA-FX SHORT TUBE, 130D, 135D and 140D, VERSA-FX STANDARD TUBE, 1 and VERSA-FX SPRCNDLR TUBE, 9 Item codes starting with 001980, 001981, 0011990 and 0011991

Reason for recall

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Data sourced from openFDA. This site is unofficial and independent of the FDA.