Recall Z-1003-2023— Siemens Medical Solutions USA, Inc

Class II · Ongoing

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2022-12-22. It names 3 FDA 510(k) clearances: K220814, K211373, K200524. Product: Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the following CT (Computed tomography systems): SOMATOM go.Now Model 11061610 NA* SOMATOM go.Now Model 11061618 NA* SOMATOM go.Up Model 11061620 SOMATOM go.Up Model 11061628 NA* SOMATOM go.All Model 11061630 SOMATOM go.All Model 11061638 NA* SOMATOM go.Top Model 11061640 SOMATOM go.Top Model 11061648 NA* SOMATOM X.cite Model 11330001 SOMATOME X.ceed Model 11330002 SOMATOM go.Sim Model 11061660 NA* SOMATOM go.Open Pro Model 11061670 NAEOTOM Alpha Model 11330003. Reason: In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped.

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Ongoing
Initiated
2022-12-22
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the following CT (Computed tomography systems): SOMATOM go.Now Model 11061610 NA* SOMATOM go.Now Model 11061618 NA* SOMATOM go.Up Model 11061620 SOMATOM go.Up Model 11061628 NA* SOMATOM go.All Model 11061630 SOMATOM go.All Model 11061638 NA* SOMATOM go.Top Model 11061640 SOMATOM go.Top Model 11061648 NA* SOMATOM X.cite Model 11330001 SOMATOME X.ceed Model 11330002 SOMATOM go.Sim Model 11061660 NA* SOMATOM go.Open Pro Model 11061670 NAEOTOM Alpha Model 11330003

Reason for recall

In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped

Data sourced from openFDA. This site is unofficial and independent of the FDA.