Recall Z-1001-2015— Genetic Testing Institute,inc

Class II · Terminated

Class II FDA medical device recall by Genetic Testing Institute,inc, initiated 2014-12-15, terminated 2015-04-30. It names one FDA 510(k) clearance: K053559. Product: PF4 Concentrated Wash (10x), REF 403624, packaged in the following assay kits: 1. ImmucorGamma LIFECODES PF4 IgG (catalog numbers HAT13G, HAT45G) and 2. ImmucorGamma LIFECODES PF4 Enhanced (catalog numbers X-HAT13, X-HAT45) The PF4 IgG assay is an Enzyme Linked Immunosorbent Assay (ELISA). The PF4 IgG ELISA is intended to detect IgG antibodies in human serum that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG kit contains all of the reagents necessary to perform the assay.. Reason: Some Lots of PF4 Concentrated Wash (10X) are a slightly more dilute concentration. This may lead to slightly increased Optical Density (OD) values. Samples with a higher background reactivity may demonstrate false positive result (borderline reactivity) at or near the cutoff for the assay..

Recalling firm
Genetic Testing Institute,inc
Classification
Class II
Status
Terminated
Initiated
2014-12-15
Terminated
2015-04-30
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

PF4 Concentrated Wash (10x), REF 403624, packaged in the following assay kits: 1. ImmucorGamma LIFECODES PF4 IgG (catalog numbers HAT13G, HAT45G) and 2. ImmucorGamma LIFECODES PF4 Enhanced (catalog numbers X-HAT13, X-HAT45) The PF4 IgG assay is an Enzyme Linked Immunosorbent Assay (ELISA). The PF4 IgG ELISA is intended to detect IgG antibodies in human serum that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG kit contains all of the reagents necessary to perform the assay.

Reason for recall

Some Lots of PF4 Concentrated Wash (10X) are a slightly more dilute concentration. This may lead to slightly increased Optical Density (OD) values. Samples with a higher background reactivity may demonstrate false positive result (borderline reactivity) at or near the cutoff for the assay.