Recall Z-1000-2022— NeuMoDx Molecular Inc
Class II · Ongoing
Class II FDA medical device recall by NeuMoDx Molecular Inc, initiated 2022-02-18. FDA's recall record names no specific 510(k) clearance for this event. Product: NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays: NeuMoDx SARS CoV-2 Test Strip (REF 300800) NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip (REF 300900) NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection Note: Results for targets other than SARS-CoV-2 are not impacted by this Urgent Medical Device Correction Ref. Number: 100100. Reason: Potential for false positive results for SARS-CoV-2 in identified lots of cartridges, when used in conjunction with assays containing SARS-CoV-2 targets (NeuMoDx" SARS-CoV-2 Test Strip Cat #300800 and NeuMoDx" Flu A-B/RSV/SARS-CoV-2 Vantage Assay Cat #300900) result incorrect management of patients.
- Recalling firm
- NeuMoDx Molecular Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-02-18
- Firm location
- Ann Arbor, MI, United States
Product description
NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays: NeuMoDx SARS CoV-2 Test Strip (REF 300800) NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip (REF 300900) NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection Note: Results for targets other than SARS-CoV-2 are not impacted by this Urgent Medical Device Correction Ref. Number: 100100
Reason for recall
Potential for false positive results for SARS-CoV-2 in identified lots of cartridges, when used in conjunction with assays containing SARS-CoV-2 targets (NeuMoDx" SARS-CoV-2 Test Strip Cat #300800 and NeuMoDx" Flu A-B/RSV/SARS-CoV-2 Vantage Assay Cat #300900) result incorrect management of patients
Data sourced from openFDA. This site is unofficial and independent of the FDA.