Recall Z-0998-2019— Ossur Americas

Class II · Terminated

Class II FDA medical device recall by Ossur Americas, initiated 2018-06-21, terminated 2020-09-04. It names one FDA 510(k) clearance: K040363. Product: Ossur Skull Pin, Sterile Ceramic Tip Pin, Part/ Description: 516CS/ Skull Pin 2-1/4in CT Sterl 1PK; 516DS/ Skull Pin 3in CT Sterl 1PK; 517CS/ Skull Pin 2-1/4in CT Sterl 4PK; 517DS/ Skull Pin 3in CT Sterl 4 PK; 522CS/ Skull Pin 2-1/4in CT Sterl 6PK; 522C/ SKULL PIN 2-1/4 CT 6 PK. Reason: Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin..

Recalling firm
Ossur Americas
Classification
Class II
Status
Terminated
Initiated
2018-06-21
Terminated
2020-09-04
Firm location
Foothill Ranch, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ossur Skull Pin, Sterile Ceramic Tip Pin, Part/ Description: 516CS/ Skull Pin 2-1/4in CT Sterl 1PK; 516DS/ Skull Pin 3in CT Sterl 1PK; 517CS/ Skull Pin 2-1/4in CT Sterl 4PK; 517DS/ Skull Pin 3in CT Sterl 4 PK; 522CS/ Skull Pin 2-1/4in CT Sterl 6PK; 522C/ SKULL PIN 2-1/4 CT 6 PK

Reason for recall

Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.

Data sourced from openFDA. This site is unofficial and independent of the FDA.