Recall Z-0994-2014— Integra LifeSciences Corp.

Class II · Terminated

Class II FDA medical device recall by Integra LifeSciences Corp., initiated 2013-12-05, terminated 2017-01-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Integra Cusa Excel Rx Only CEM" Nosecone for CUSA¿ Excel System Product Usage: Intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator The CUSA Electrosurgical Module (CEM) provides Desiccate Coagulation waveform electrosurgical capability to CUSA Excel handpieces. The CUSA handpiece with the CEM Nosecone works in conjunction with the Force FX generator in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.. Reason: The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position..

Recalling firm
Integra LifeSciences Corp.
Classification
Class II
Status
Terminated
Initiated
2013-12-05
Terminated
2017-01-26
Firm location
Plainsboro, NJ, United States

Product description

Integra Cusa Excel Rx Only CEM" Nosecone for CUSA¿ Excel System Product Usage: Intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator The CUSA Electrosurgical Module (CEM) provides Desiccate Coagulation waveform electrosurgical capability to CUSA Excel handpieces. The CUSA handpiece with the CEM Nosecone works in conjunction with the Force FX generator in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.

Reason for recall

The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position.

Data sourced from openFDA. This site is unofficial and independent of the FDA.