Recall Z-0991-2022— Baxter Healthcare Corporation
Class II · Ongoing
Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2022-03-15. FDA's recall record names no specific 510(k) clearance for this event. Product: PrisMax System, Product Code 955626. Reason: This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-03-15
- Firm location
- Deerfield, IL, United States
Product description
PrisMax System, Product Code 955626
Reason for recall
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
Data sourced from openFDA. This site is unofficial and independent of the FDA.