Recall Z-0990-2022— St. Jude Medical, Cardiac Rhythm Management Division

Class II · Ongoing

Class II FDA medical device recall by St. Jude Medical, Cardiac Rhythm Management Division, initiated 2022-03-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Gallant HF Implantable Cardioverter Defibrillator REF# CDHFA500Q. Reason: Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity..

Recalling firm
St. Jude Medical, Cardiac Rhythm Management Division
Classification
Class II
Status
Ongoing
Initiated
2022-03-10
Firm location
Sylmar, CA, United States

Product description

Gallant HF Implantable Cardioverter Defibrillator REF# CDHFA500Q

Reason for recall

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Data sourced from openFDA. This site is unofficial and independent of the FDA.