Recall Z-0990-2013— Biomerieux Inc

Class II · Terminated

Class II FDA medical device recall by Biomerieux Inc, initiated 2013-02-18, terminated 2014-08-14. FDA's recall record names no specific 510(k) clearance for this event. Product: VITEK 2 XL computer system when equipped with the following Hewlett-Packard PC models: (1) rp5700, Cat. #W0452; (2) rp5700 (refurbished), Cat. #W0452R; (3) rp5800 XPE, Cat. 413642; (4) rp5800 WES7, Cat. #413862; (5) dc7100 8-port, Cat. #W0441; (6) dc7100 8-port (refurbished), Cat. #W0441R; (7) dc7700 8-port, Cat. #W0447; (8) dc7700 8-port (refurbished), Cat. #W0447R; (9) dc7800, Cat. #W0449; and (10) dc7800 (refurbished), Cat. #W0449R. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.. Reason: There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS..

Recalling firm
Biomerieux Inc
Classification
Class II
Status
Terminated
Initiated
2013-02-18
Terminated
2014-08-14
Firm location
Hazelwood, MO, United States

Product description

VITEK 2 XL computer system when equipped with the following Hewlett-Packard PC models: (1) rp5700, Cat. #W0452; (2) rp5700 (refurbished), Cat. #W0452R; (3) rp5800 XPE, Cat. 413642; (4) rp5800 WES7, Cat. #413862; (5) dc7100 8-port, Cat. #W0441; (6) dc7100 8-port (refurbished), Cat. #W0441R; (7) dc7700 8-port, Cat. #W0447; (8) dc7700 8-port (refurbished), Cat. #W0447R; (9) dc7800, Cat. #W0449; and (10) dc7800 (refurbished), Cat. #W0449R. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.

Reason for recall

There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Data sourced from openFDA. This site is unofficial and independent of the FDA.