Recall Z-0989-2022— St. Jude Medical, Cardiac Rhythm Management Division
Class II · Ongoing
Class II FDA medical device recall by St. Jude Medical, Cardiac Rhythm Management Division, initiated 2022-03-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Gallant VR Implantable Cardioverter Defibrillator REF # CDVRA500Q. Reason: Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity..
- Recalling firm
- St. Jude Medical, Cardiac Rhythm Management Division
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-03-10
- Firm location
- Sylmar, CA, United States
Product description
Gallant VR Implantable Cardioverter Defibrillator REF # CDVRA500Q
Reason for recall
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.