Recall Z-0988-2025— St. Jude Medical

Class II · Ongoing

Class II FDA medical device recall by St. Jude Medical, initiated 2024-10-05. FDA's recall record names no specific 510(k) clearance for this event. Product: CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000. Reason: As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings..

Recalling firm
St. Jude Medical
Classification
Class II
Status
Ongoing
Initiated
2024-10-05
Firm location
Atlanta, GA, United States

Product description

CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000

Reason for recall

As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.

Data sourced from openFDA. This site is unofficial and independent of the FDA.