Recall Z-0980-2022— Siemens Medical Solutions USA, Inc
Class II · Ongoing
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2022-02-23. It names one FDA 510(k) clearance: K181407. Product: Artis Q.zen biplane, Model Number 10848355. Reason: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT"..
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-02-23
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Artis Q.zen biplane, Model Number 10848355
Reason for recall
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Data sourced from openFDA. This site is unofficial and independent of the FDA.