Recall Z-0975-2022— Siemens Medical Solutions USA, Inc

Class II · Ongoing

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2022-02-23. It names one FDA 510(k) clearance: K181407. Product: Artis Q ceiling, Model Number 10848281. Reason: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT"..

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Ongoing
Initiated
2022-02-23
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Artis Q ceiling, Model Number 10848281

Reason for recall

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Data sourced from openFDA. This site is unofficial and independent of the FDA.