Recall Z-0972-2026— Zimmer, Inc.

Class II · Ongoing

Class II FDA medical device recall by Zimmer, Inc., initiated 2025-12-02. It names one FDA 510(k) clearance: K241651. Product: Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509420 Software Version: N/A Product Description: Affixus 9 mm, Length 420 mm, Right, Trochanteric Nail Component: N/A. Reason: Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-12-02
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509420 Software Version: N/A Product Description: Affixus 9 mm, Length 420 mm, Right, Trochanteric Nail Component: N/A

Reason for recall

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Data sourced from openFDA. This site is unofficial and independent of the FDA.