Recall Z-0972-2017— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2016-12-08, terminated 2019-04-09. It names one FDA 510(k) clearance: K123529. Product: Artis Q/Q.zen, Angiographic x-ray system Stand alone system The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee/ zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee/ zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis. Reason: Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: - In Artis Systems with A100 Plus or A100G generators, an attempt to resume operation following detection of a fault can result in the failure of a module in the high-voltage generator. - For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with graphics card may, in rare cases, result in the loss of image display in the examination room..

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2016-12-08
Terminated
2019-04-09
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Artis Q/Q.zen, Angiographic x-ray system Stand alone system The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee/ zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee/ zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis

Reason for recall

Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: - In Artis Systems with A100 Plus or A100G generators, an attempt to resume operation following detection of a fault can result in the failure of a module in the high-voltage generator. - For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with graphics card may, in rare cases, result in the loss of image display in the examination room.