Recall Z-0970-2019— PENTAX of America Inc

Class II · Terminated

Class II FDA medical device recall by PENTAX of America Inc, initiated 2018-12-03, terminated 2020-08-21. It names one FDA 510(k) clearance: K163684. Product: C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus.. Reason: The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur..

Recalling firm
PENTAX of America Inc
Classification
Class II
Status
Terminated
Initiated
2018-12-03
Terminated
2020-08-21
Firm location
Redwood City, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

C2 CryoBalloon Controller, REF: FG-1017, with Controller Software v1.18.258 The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus.

Reason for recall

The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur.