Recall Z-0965-2022— St. Jude Medical, Cardiac Rhythm Management Division

Class II · Ongoing

Class II FDA medical device recall by St. Jude Medical, Cardiac Rhythm Management Division, initiated 2022-03-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Merlin 2 PCS MER3700 programmer Model MER3400 software. Reason: Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters..

Recalling firm
St. Jude Medical, Cardiac Rhythm Management Division
Classification
Class II
Status
Ongoing
Initiated
2022-03-10
Firm location
Sylmar, CA, United States

Product description

Merlin 2 PCS MER3700 programmer Model MER3400 software

Reason for recall

Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.

Data sourced from openFDA. This site is unofficial and independent of the FDA.