Recall Z-0965-2014— Bio-Rad Laboratories, Inc.
Class III · Terminated
Class III FDA medical device recall by Bio-Rad Laboratories, Inc., initiated 2013-11-22, terminated 2014-03-20. It names one FDA 510(k) clearance: K031043. Product: D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.. Reason: D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System..
- Recalling firm
- Bio-Rad Laboratories, Inc.
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2013-11-22
- Terminated
- 2014-03-20
- Firm location
- Hercules, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.
Reason for recall
D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System.