Recall Z-0961-2024— Wipro GE Healthcare Private Ltd.

Class I · Ongoing

Class I FDA medical device recall by Wipro GE Healthcare Private Ltd., initiated 2023-12-26. FDA's recall record names no specific 510(k) clearance for this event. Product: GE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659, M1206259, M1206260, M1166836, M1166814. Reason: GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling..

Recalling firm
Wipro GE Healthcare Private Ltd.
Classification
Class I
Status
Ongoing
Initiated
2023-12-26
Firm location
Bengaluru, N/A, India

Product description

GE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659, M1206259, M1206260, M1166836, M1166814

Reason for recall

GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

Data sourced from openFDA. This site is unofficial and independent of the FDA.