Recall Z-0960-2024— Wipro GE Healthcare Private Ltd.
Class I · Ongoing
Class I FDA medical device recall by Wipro GE Healthcare Private Ltd., initiated 2023-12-26. FDA's recall record names no specific 510(k) clearance for this event. Product: GE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, and 6600-0970-500. Reason: GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling..
- Recalling firm
- Wipro GE Healthcare Private Ltd.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-12-26
- Firm location
- Bengaluru, N/A, India
Product description
GE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, and 6600-0970-500
Reason for recall
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
Data sourced from openFDA. This site is unofficial and independent of the FDA.