Recall Z-0959-2021— Prytime Medical Devices, Inc.

Class II · Terminated

Class II FDA medical device recall by Prytime Medical Devices, Inc., initiated 2020-12-14, terminated 2022-09-28. FDA's recall record names no specific 510(k) clearance for this event. Product: PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC, STERILE, 1 EA, SYRINGE, 30CC, STERILE 1 EA, INTRODUCER SHEATH KIT, 7 FR W/.035 SHORT GUIDEWIRE, STERILE 1 EA., PREFILLED, STERILE FIELD SALINE SYRINGE, 10 ML, 4 EA, CLAMP, 5 FR, STERILE, 1 EA, SCALPEL, #11SS, STERILE, 1 EA, SUTURE, NYLON, 18", 2-0, REVERSE CUTTING, STERILE 1 EA, THREE QUARTER DRAPE, STERILE 1 EA, INSERT, SALINE 1 EA, INSERT, CLAMP 1 EA UDI: (01) 0 0863092 00010 1. Reason: There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G needle. If the 21G needle is used to gain vascular access then the guidewire included in the kit will not pass through the access needle to facilitate insertion of an introducer sheath and could therefore cause procedure delay or injury..

Recalling firm
Prytime Medical Devices, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-12-14
Terminated
2022-09-28
Firm location
Boerne, TX, United States

Product description

PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC, STERILE, 1 EA, SYRINGE, 30CC, STERILE 1 EA, INTRODUCER SHEATH KIT, 7 FR W/.035 SHORT GUIDEWIRE, STERILE 1 EA., PREFILLED, STERILE FIELD SALINE SYRINGE, 10 ML, 4 EA, CLAMP, 5 FR, STERILE, 1 EA, SCALPEL, #11SS, STERILE, 1 EA, SUTURE, NYLON, 18", 2-0, REVERSE CUTTING, STERILE 1 EA, THREE QUARTER DRAPE, STERILE 1 EA, INSERT, SALINE 1 EA, INSERT, CLAMP 1 EA UDI: (01) 0 0863092 00010 1

Reason for recall

There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G needle. If the 21G needle is used to gain vascular access then the guidewire included in the kit will not pass through the access needle to facilitate insertion of an introducer sheath and could therefore cause procedure delay or injury.

Data sourced from openFDA. This site is unofficial and independent of the FDA.