Recall Z-0958-2026— Medtronic MiniMed, Inc.

Class II · Ongoing

Class II FDA medical device recall by Medtronic MiniMed, Inc., initiated 2025-11-13. It names one FDA 510(k) clearance: K242775. Product: InPen App, Model/CFN Number: MMT-8061 (Android Users). Reason: When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia..

Recalling firm
Medtronic MiniMed, Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-11-13
Firm location
Northridge, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

InPen App, Model/CFN Number: MMT-8061 (Android Users)

Reason for recall

When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.

Data sourced from openFDA. This site is unofficial and independent of the FDA.