Recall Z-0957-2016— Xeridiem Mediem Medical Devices Inc
Class II · Terminated
Class II FDA medical device recall by Xeridiem Mediem Medical Devices Inc, initiated 2015-12-23, terminated 2016-10-25. It names one FDA 510(k) clearance: K142297. Product: EndoVive 3s Low Profile Balloon Kits Part Number: M00548720 (XMD P/N 70-0050-718) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.. Reason: A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances..
- Recalling firm
- Xeridiem Mediem Medical Devices Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-12-23
- Terminated
- 2016-10-25
- Firm location
- Tucson, AZ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EndoVive 3s Low Profile Balloon Kits Part Number: M00548720 (XMD P/N 70-0050-718) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.
Reason for recall
A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.