Recall Z-0956-2022— Abbott
Class II · Ongoing
Class II FDA medical device recall by Abbott, initiated 2022-03-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve D-F, Model A-TCSE-DF, Sterile.. Reason: Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device..
- Recalling firm
- Abbott
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-03-24
- Firm location
- Plymouth, MN, United States
Product description
Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve D-F, Model A-TCSE-DF, Sterile.
Reason for recall
Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.
Data sourced from openFDA. This site is unofficial and independent of the FDA.